EFD-0392 FieldMan E-Field probe, 100kHz - 3GHz
  • EFD-0392 FieldMan E-Field probe, 100kHz - 3GHz

EFD-0392 FieldMan E-Field probe, 100kHz - 3GHz

EFD-0392

Isotropic measurement of electric fields from 100kHz to 3GHz using the Field Meter FieldMan®

Description

The probe detects electric fields from 100 kHz to 3 GHz, covering the fields that occur in broadcasting, telecoms, and industry. The dynamic range from 0.5 V/m up to 1,300 V/m (68 dB) makes the probe ideal for measuring exposure in both the general public and the occupational environment. The probe is characterized by particularly high power handling and a wide True RMS range.

The probe's interface digitally transmits the measurement data to the base unit, which has no individual influence on the measured values and therefore does not need to be calibrated. The accredited probe calibration is carried out at several frequencies. The calibration data is stored in the probe and is automatically taken into account during the measurement. If the frequency of the predominating field strength is known, a correction factor can additionally be applied to increase the measurement accuracy.

  • General public and occupational field exposure from broadcasting, telecoms and industrial equipment
  • Isotropic (non-directional) measurement
  • 68 dB dynamic without changing range
  • For high field strengths up to 1300 V/m
  • Digital probe interface - no more meter calibration
  • Self-test of the probe interface with integrated sensor function test
  • Automatic offset correction, no zero adjustment required
  • Temperature compensation of the sensors for minimum drift and wide temperature range down to -20 °C
  • High Immunity at 50/60 Hz
  • Accredited calibration included
Product Details
EFD-0392

Data sheet

Field
E-Field
Dimensions
308 mm x 66 mm Ø
Weight
up to 100 g
Frequency range
100kHz - 3GHz
Operating conditions
-20 °C to +50 °C, less than 29 g/m³ (less than 93 % RH at +30 °C), non-condensing
Storage conditions
1K5 (IEC 60721-3) -40 °C to +70 °C
Protection class
IP54 (probe screwed on)
Type of product
FieldMan probes
Dynamic range (dB)
68 dB
Measurement range, E-field
0.5 to 1300 V/m (CW); 0.5 to 36 V/m (True RMS); 66 nW/cm² to 450 mW/cm² (CW); 66 nW/cm² to 0.35 mW/cm² (True RMS)
Integration time
265 ms
Type of frequency response
Flat
Continuous wave damage level
2000 V/m; 1000 mW/cm²
Peak damage level
20 kV/m; 100 W/cm²
Sensor type
Diode based system
Directivity
Isotropic (Tri-axial)
Spatial assessment
3 separate axes
Sampling rate
5 Hz
Temperature sensors
Integrated sensors for displaying the ambient temperature, for automatic offset compensation and for compensation of the temperature response
Self-test
Interface function test and sensor test for short circuit and interruption of diodes
Flatness of frequency response
±1 dB (1 MHz to 300 MHz); +1.4/ -1 dB (300 MHz to 2.45 GHz)
Linearity deviation
+2/-3 dB (1 to 2 V/m); ±1 dB (2 to 4 V/m); ±0.5 dB (4 to 400 V/m); ±1 dB (400 to 1300 V/m). +2/-3 dB (0.265 to 1.06 μW/cm²); ±1 dB (1.06 to 4.25 μW/cm²); ±0.5 dB (4.25 μW/cm² to 42 mW/cm²); ±1 dB (42 to 450 mW/cm²)
Isotropic deviation
±1 dB
Temperature response
+0.1/ -0.8 dB (0 °C to 50 °C, related to 23 °C); +1.5/ -0.8 dB (-20 °C to +50 °C, related to 23 °C)
Accredited calibration
DAkkS, ILAC-MRA (DIN EN ISO/IEC 17025, IEEE Std. 1309). For measurands outside the scope, a factory calibration is performed.
Recommended calibration interval
24 months
Attachments
Download
Datasheet Probe EFD-0392

Datasheet Probe EFD-0392, v. NSTS 1122-E0386A, English

Download (307.99KB)
Comments (0)
Grade
No customer reviews for the moment.
10 other products in the same category:
You might also like

Reference: 2460/101

Brand: Narda Safety Test Solutions

Electro and magnetic field meter FieldMan

(0)
The electromagnetic field meter and successor of the NBM series ranging from 0 Hz to 90 GHz. Narda has opened up new dimensions in the versatility and handling of electromagnetic field measuring devices with the FieldMan. This light, easy to use instrument can be operated with various probes (also NBM-probes) to provide reliable, omnidirectional...