EFD-9091 FieldMan E-Field probe, 100MHz - 90GHz
  • EFD-9091 FieldMan E-Field probe, 100MHz - 90GHz

EFD-9091 FieldMan E-Field probe, 100MHz - 90GHz

EFD-9091

Isotropic measurement of electric fields from 100 MHz to 90 GHz using the Field Meter FieldMan®

Description

The probe detects electric fields from 100 MHz up to 90 GHz. This area covers the frequencies used for radio link, as well as applications for telecommunications, satellite communications and radar technologies. The probe’s high linearity and sensitivity make it ideal for checking human safety limit values in the general public and occupational environments.

The probe's interface digitally transmits the measurement data to the base unit, which has no individual influence on the measured values and therefore does not need to be calibrated. Probe calibration is carried out at several frequencies and is accredited up to 18 GHz. The calibration data is stored in the probe and is automatically taken into account during the measurement. If the frequency of the predominating field strength is known, a correction factor can additionally be applied to increase the measurement accuracy.

  • Isotropic (non-directional) measurement
  • Excellent sensitivity starting at 0.5 V/m
  • Wide True RMS range up to 61.4 V/m
  • Digital probe interface - no more meter calibration
  • Self-test of the probe interface with integrated sensor function test
  • Automatic offset correction, no zero adjustment required
  • Wide temperature range from -20 °C to +50 °C
Product Details
EFD-9091

Data sheet

Field
E-Field
Dimensions
308 mm x 66 mm Ø
Weight
up to 100 g
Frequency range
100MHz - 90GHz
Operating conditions
-20 °C to +50 °C
up to 29 g/m³ (up to 93 % RH at +30 °C), non-condensing
Protection class
IP54 (probe screwed on)
Type of product
FieldMan probes
Dynamic range (dB)
58
Measurement range, E-field
0.5 to 400 V/m (CW), 66 nW/cm² to 42 mW/cm² (CW)
Typ. 0.5 to 61.4 V/m (True RMS), 2 V/m for frequencies more than 60 GHz
Type of frequency response
Flat
Continuous wave damage level
1600 V/m, 700 mW/cm²
Peak damage level
1900 V/m, 1 W/cm²
Sensor type
Diode based system
Directivity
Isotropic (Tri-axial)
Spatial assessment
Combined 3-axis (RSS)
Sampling rate
5 Hz / 275 ms
Temperature sensors
Integrated sensors for displaying the ambient temperature, for automatic offset compensation and for compensation of the temperature response
Self-test
Interface function test and sensor test for interruption of diodes
Flatness of frequency response
+3/-6 dB (100 MHz to 60 GHz)
typ. +3/-10 dB (from 60 GHz to 90 GHz)
±3 dB (300 MHz to 40 GHz)
Linearity deviation
±1 dB (2 to 250 V/m)
±1 dB (from 1 μW/cm² to 16.5 mW/cm²)
±2 dB (250 to 400 V/m)
±2 dB (from 16.5 mW/cm² to 42.44 mW/cm²)
±2 dB (from 16.5 mW/cm² to 170 mW/cm²)
±3 dB (0.265 μW/cm² to 1 μW/cm²)
±3 dB (1 to 2 V/m)
Isotropic deviation
±1.25 dB (100 MHz to 10 GHz)
±2 dB (more than 10 GHz to 26.5 GHz)
±2 dB typ. (more than 26.5 GHz)
Temperature response
±0.2 dB (-20 °C to +50 °C, related to 23 °C)
Accredited calibration
DAkkS, ILAC-MRA (DIN EN ISO/IEC 17025, IEEE Std. 1309)
For measurands outside the scope, a factory calibration is performed.
Recommended calibration interval
24 months
Attachments
Download
FieldMan Brochure

FieldMan Brochure, v. NSTS 03/23 E0384B, English

Download (6.51MB)
FieldMan Sondes

FieldMan - Probes program for measurements up to 90 GHz

Download (100.62KB)
Datasheet Probe EFD-9091

Datasheet Probe EFD-9091, v. NSTS 0523-E0405A, English

Download (220.38KB)
Comments (0)
Grade
No customer reviews for the moment.
10 other products in the same category:
You might also like

Reference: 2460/101

Brand: Narda Safety Test Solutions

Electro and magnetic field meter FieldMan

(0)
The electromagnetic field meter and successor of the NBM series ranging from 0 Hz to 90 GHz. Narda has opened up new dimensions in the versatility and handling of electromagnetic field measuring devices with the FieldMan. This light, easy to use instrument can be operated with various probes (also NBM-probes) to provide reliable, omnidirectional...