EFD-4091 FieldMan E-Field probe, 2MHz - 40GHz
  • EFD-4091 FieldMan E-Field probe, 2MHz - 40GHz

EFD-4091 FieldMan E-Field probe, 2MHz - 40GHz

EFD-4091

Isotropic measurement of electric fields from 2 MHz to 40 GHz using the Field Meter FieldMan®

Kuvaus

The probe detects electric fields from 2 MHz up to 40 GHz. This area covers the frequencies used for broadcasting, as well as applications for telecommunications, satellite communications and radar technologies. The probe’s high linearity and sensitivity make it ideal for checking human safety limit values in the general public and occupational environments. 

The probe's interface digitally transmits the measurement data to the base unit, which has no individual influence on the measured values and therefore does not need to be calibrated. Probe calibration is carried out at several frequencies and is accredited up to 18 GHz. The calibration data is stored in the probe and is automatically taken into account during the measurement. If the frequency of the predominating field strength is known, a correction factor can additionally be applied to increase the measurement accuracy.

  • Isotropic (non-directional) measurement
  • Excellent sensitivity starting at 1.2 V/m
  • Wide True RMS range up to 65 V/m
  • Digital probe interface - no more meter calibration
  • Self-test of the probe interface with integrated sensor function test
  • Automatic offset correction, no zero adjustment required
  • Wide temperature range from -20 °C to +50 °C
  • Calibration is performed up to 40 GHz and is accredited up to 18 GHz
Tuotetiedot
EFD-4091

Lisätiedot

Kenttä
Sähkökenttä
Mitat
308 mm x 66 mm Ø
Paino
up to 100 g
Taajuusalue
2MHz - 40GHz
Käyttöolosuhtee
-20 °C to +50 °C
up to 29 g/m³ (up to 93 % RH at +30 °C), non-condensing
Suojausluokka
IP54 (probe screwed on)
Tuotteen tyyppi
FieldMan probes
Dynaaminen alue (dB)
56.5
Mittausalue, E-kenttä
1.2 to 65 V/m (True RMS), 0.382 μW/cm² to 1.0 mW/cm² (True RMS)
1.2 to 800 V/m (CW), 0.382 μW/cm² to 170 mW/cm² (CW)
Type of frequency response
Flat
Jatkuva aaltovauriotaso
1500 V/m, 600 mW/cm²
Huippuvaurio
1800 V/m, 860 mW/cm²
Anturin tyyppi
Diode based system
Suuntaus
Isotrooppinen (kolmiakselinen)
Tila-arviointi
Combined 3-axis (RSS)
Näytteenottotaajuus
5 Hz / 265 ms
Lämpötila-anturit
Integrated sensors for displaying the ambient temperature, for automatic offset compensation and for compensation of the temperature response
Itsetestaus
Interface function test and sensor test for interruption of diodes
Taajuusvasteen tasaisuus
typ. -3dB @ 2 MHz / 40 GHz
±1.5 dB (3 MHz to 15 GHz)
±3 dB (from 15 GHz to 35 GHz)
Lineaarisuuden poikkeama
±1 dB (1 μW/cm² to 16.5 mW/cm²)
±1 dB (2 to 250 V/m)
±1 dB (from 1 μW/cm² to 16.5 mW/cm²)
±2 dB (from 16.5 mW/cm² to 170 mW/cm²)
±2 dB (from 250 to 800 V/m)
±3 dB (0.38 to 1 μW/cm²)
±3 dB (1 to 2 V/m)
Isotrooppinen poikkeama
typ. ±2 dB @ 2 MHz
typ. ±6 dB @ 40 GHz
±1.0 dB (10 MHz to 5 GHz)
±1.5 dB (from 5 GHz to 20 GHz)
±2 dB (more than 10 GHz to 26.5 GHz)
±3.0 dB (from 20 GHz to 30 GHz)
Lämpötilavaste
+0.5/ -0.5 dB (0 °C to 50 °C, related to 23 °C)
+1.5/ -1.5 dB (-20 °C to +50 °C, related to 23 °C)
Akkreditoitu kalibrointi
DAkkS, ILAC-MRA (DIN EN ISO/IEC 17025, IEEE Std. 1309)
For measurands outside the scope, a factory calibration is performed.
Suositeltu kalibrointiväli
24 kuukautta
Liitteet
Lataa
FieldMan Brochure

FieldMan Brochure, v. NSTS 03/23 E0384B, English

Lataa (6.51MB)
FieldMan Sondes

FieldMan - Probes program for measurements up to 90 GHz

Lataa (100.62KB)
Datasheet Probe EFD-4091

Datasheet Probe EFD-4091, v. NSTS 0523-E0399A, English

Lataa (229.55KB)
Kommentit (0)
Arvosana
Tällä hetkellä ei ole asiakkaiden arvosteluja.
10 muuta tuotetta samassa kategoriassa:
Saatat pitää myös

Viite: 2460/101

Tuotemerkki: Narda Safety Test Solutions

Electro and magnetic field meter FieldMan

(0)
The electromagnetic field meter and successor of the NBM series ranging from 0 Hz to 90 GHz. Narda has opened up new dimensions in the versatility and handling of electromagnetic field measuring devices with the FieldMan. This light, easy to use instrument can be operated with various probes (also NBM-probes) to provide reliable, omnidirectional...