Isotropic measurement of electric fields from 300 MHz to 50 GHz using the Field Meter FieldMan®

Aprašymas

The probe contains three orthogonally-arranged thermo-couple sensors and detects electric fields from 300 MHz to 50 GHz, covering the fields found in telecommunications, satellite communications and radar equipment. The use of thermocouples naturally results in a true RMS reading, making the probe particularly suitable for measuring human safety limit values in a multi-frequency environment.

The probe's interface digitally transmits the measurement data to the base unit, which has no individual influence on the measured values and therefore does not need to be calibrated. The accredited probe calibration is carried out at several frequencies. The calibration data is stored in the probe and is automatically taken into account during the measurement. If the frequency of the predominating field strength is known, a correction factor can additionally be applied to increase the measurement accuracy.

  • Measurement of field exposure from telecommunications, satellite communications and radar in working environments
  • True RMS value display even when several, strongly pulsed signals are superimposed
  • Isotropic (non-directional) measurement
  • Digital probe interface - no more meter calibration
  • Self-test of the probe interface with integrated sensor function test
  • Automatic offset correction, no zero adjustment required
  • Lowest temperature response due to thermocouple sensor technology and wide temperature range down to -20 °C
  • High Immunity at 50/60 Hz
  • Accredited calibration included
Prekė detaliau
EFD-5091

Charakteristikos

Laukas
Elektrinis laukas
Matmenys
308 mm x 66 mm Ø
Svoris
up to 100 g
Dažnių diapazonas
300MHz - 50GHz
Eksploatacijos sąlygos
-20 °C to +50 °C; up to 29 g/m³ (up to 93 % RH at +30 °C), non-condensing
Laikymo sąlygos
1K5 (IEC 60721-3) -40 °C to +70 °C
Apsaugos klasė
IP54 (probe screwed on)
Produkto tipas
FieldMan probes
Dinaminis diapazonas (dB)
37 dB
Matavimo diapazonas, E-laukas
8 to 614 V/m (True RMS); 17 μW/cm² to 100 mW/cm² (True RMS)
Integracijos laikas
270 ms
Type of frequency response
Flat
Nuolatinės bangos pažeidimo lygis
1200 V/m; 0.4 W/cm²
Didžiausia žala
20 kV/m; 100 W/cm²
Jutiklio tipas
Thermocouple
Kryptingumas
Izotropinis (triašis)
Erdvinis vertinimas
Combined 3-axis (RSS)
Atrankos dažnis
5 Hz
Temperatūros jutikliai
Integrated sensors for displaying the ambient temperature and for automatic offset compensation
Savęs išbandymas
Interface function test and sensor test for interruption of the sensors
Dažnio atsako plokštumas
±1.5 dB (1.8 GHz to 26.5 GHz); +1.5/-2 dB (up to 26.5 GHz to 40 GHz)
Tiesiškumo nuokrypis
±1 dB (19.5 to 61.4 V/m); ±0.3 dB (less than 61.4 V/m); ±1 dB (0.1 to 1 mW/cm²) ±0.3 dB (less than 1 mW/cm²)
Izotropinis nuokrypis
±1 dB (0.6 GHz to 10 GHz); ±1.2 dB (10 GHz to 45.5 GHz)
Temperatūros reakcija
±0 dB
Akredituotas kalibravimas
DAkkS, ILAC-MRA (DIN EN ISO/IEC 17025, IEEE Std. 1309); For measurands outside the scope, a factory calibration is performed.
Rekomenduojamas kalibravimo intervalas
24 months
Priedai
Atsisiųsti
FieldMan Brochure