EFD-0391 FieldMan E-Field probe, 100kHz - 6GHz
  • EFD-0391 FieldMan E-Field probe, 100kHz - 6GHz

EFD-0391 FieldMan E-Field probe, 100kHz - 3GHz

EFD-0391
Narda Safety Test Solutions

Isotropic measurement of electric fields from 100kHz to 6GHz using the Field Meter FieldMan®

Aprašymas

The probe detects electric fields from 100 kHz to 3 GHz, covering the fields that occur in broadcasting, telecoms, and industry. The high sensitivity of 0.2 V/m and excellent linearity make it ideal for measuring human safety limit values in the general public domain.

The probe's interface digitally transmits the measurement data to the base unit, which has no individual influence on the measured values and therefore does not need to be calibrated. The accredited probe calibration is carried out at several frequencies. The calibration data is stored in the probe and is automatically taken into account during the measurement. If the frequency of the predominating field strength is known, a correction factor can additionally be applied to increase the measurement accuracy.

  • General public and occupational field exposure from broadcasting, telecoms and industrial equipment
  • Isotropic (non-directional) measurement
  • 64 dB dynamic without changing range
  • Excellent sensitivity starting at 0.2 V/m
  • Digital probe interface - no more meter calibration
  • Self-test of the probe interface with integrated sensor function test
  • Automatic offset correction, no zero adjustment required
  • Temperature compensation of the sensors for minimum drift and wide temperature range down to -20 °C
  • High Immunity at 50/60 Hz
  • Accredited calibration included
Prekė detaliau
EFD-0391

Charakteristikos

Laukas
Elektrinis laukas
Matmenys
308 mm x 66 mm Ø
Svoris
up to 100 g
Dažnių diapazonas
100kHz - 3GHz
Eksploatacijos sąlygos
-20 °C to +50 °C; up to 29 g/m³ (up to 93 % RH at +30 °C), non-condensing
Laikymo sąlygos
1K5 (IEC 60721-3) -40 °C to +70 °C
Apsaugos klasė
IP54 (probe screwed on)
Produkto tipas
FieldMan probes
Dinaminis diapazonas (dB)
64dB
Matavimo diapazonas, E-laukas
0.2 to 320 V/m (CW), 0.2 to 10 V/m (True RMS), 10 nW/cm² to 27 mW/cm² (CW), 10 nW/cm² to 0.027 mW/cm² (True RMS)
Integracijos laikas
265 ms
Type of frequency response
Flat
Nuolatinės bangos pažeidimo lygis
800 V/m, 170 mW/cm²
Didžiausia žala
8 kV/m, 17 W/cm²
Jutiklio tipas
Diode based system
Kryptingumas
Izotropinis (triašis)
Erdvinis vertinimas
3 separate axes
Atrankos dažnis
5 Hz
Temperatūros jutikliai
Integrated sensors for displaying the ambient temperature, for automatic offset compensation and for compensation of the temperature response
Savęs išbandymas
Interface function test and sensor test for short circuit and interruption of diodes
Dažnio atsako plokštumas
±1 dB (1 MHz to 1 GHz); ±1.25 dB (1 GHz to 2.45 GHz)
Tiesiškumo nuokrypis
±0.5 dB (1.2 to 200 V/m); ±0.7 dB (200 to 320 V/m); ±0.5 dB (0.00038 to 10.6 mW/cm²);±0.7 dB (10.6 to 27 mW/cm²)
Izotropinis nuokrypis
±1 dB
Temperatūros reakcija
+0.1/ -0.8 dB (0 °C to 50 °C, related to 23 °C); +1.5/ -0.8 dB (-20 °C to +50 °C, related to 23 °C)
Akredituotas kalibravimas
DAkkS, ILAC-MRA (DIN EN ISO/IEC 17025, IEEE Std. 1309); For measurands outside the scope, a factory calibration is performed.
Rekomenduojamas kalibravimo intervalas
24 months
Priedai
Atsisiųsti
FieldMan Brochure

FieldMan Brochure, v. NSTS 03/23 E0384B, English

Atsisiųsti (6.51MB)
Datasheet Probe EFD-0391

Datasheet Probe EFD-0391, v. NSTS 1122-E0385A, English

Atsisiųsti (311.48KB)
Komentarai (0)
Įvertinimas
Šiuo metu klientų atsiliepimų nėra.
10 kitos prekės toje pačioje kategorijoje:
Jums taip pat gali patikti