EFD-9091 FieldMan E-Field probe, 100MHz - 90GHz
  • EFD-9091 FieldMan E-Field probe, 100MHz - 90GHz

EFD-9091 FieldMan E-Field probe, 100MHz - 90GHz

EFD-9091
Narda Safety Test Solutions

Isotropic measurement of electric fields from 100 MHz to 90 GHz using the Field Meter FieldMan®

Aprašymas

The probe detects electric fields from 100 MHz up to 90 GHz. This area covers the frequencies used for radio link, as well as applications for telecommunications, satellite communications and radar technologies. The probe’s high linearity and sensitivity make it ideal for checking human safety limit values in the general public and occupational environments.

The probe's interface digitally transmits the measurement data to the base unit, which has no individual influence on the measured values and therefore does not need to be calibrated. Probe calibration is carried out at several frequencies and is accredited up to 18 GHz. The calibration data is stored in the probe and is automatically taken into account during the measurement. If the frequency of the predominating field strength is known, a correction factor can additionally be applied to increase the measurement accuracy.

  • Isotropic (non-directional) measurement
  • Excellent sensitivity starting at 0.5 V/m
  • Wide True RMS range up to 61.4 V/m
  • Digital probe interface - no more meter calibration
  • Self-test of the probe interface with integrated sensor function test
  • Automatic offset correction, no zero adjustment required
  • Wide temperature range from -20 °C to +50 °C
Prekė detaliau
EFD-9091

Charakteristikos

Laukas
Elektrinis laukas
Matmenys
308 mm x 66 mm Ø
Svoris
up to 100 g
Dažnių diapazonas
100MHz - 90GHz
Eksploatacijos sąlygos
-20 °C to +50 °C
up to 29 g/m³ (up to 93 % RH at +30 °C), non-condensing
Apsaugos klasė
IP54 (probe screwed on)
Produkto tipas
FieldMan probes
Dinaminis diapazonas (dB)
58
Matavimo diapazonas, E-laukas
0.5 to 400 V/m (CW), 66 nW/cm² to 42 mW/cm² (CW)
Typ. 0.5 to 61.4 V/m (True RMS), 2 V/m for frequencies more than 60 GHz
Type of frequency response
Flat
Nuolatinės bangos pažeidimo lygis
1600 V/m, 700 mW/cm²
Didžiausia žala
1900 V/m, 1 W/cm²
Jutiklio tipas
Diode based system
Kryptingumas
Izotropinis (triašis)
Erdvinis vertinimas
Combined 3-axis (RSS)
Atrankos dažnis
5 Hz / 275 ms
Temperatūros jutikliai
Integrated sensors for displaying the ambient temperature, for automatic offset compensation and for compensation of the temperature response
Savęs išbandymas
Interface function test and sensor test for interruption of diodes
Dažnio atsako plokštumas
+3/-6 dB (100 MHz to 60 GHz)
typ. +3/-10 dB (from 60 GHz to 90 GHz)
±3 dB (300 MHz to 40 GHz)
Tiesiškumo nuokrypis
±1 dB (2 to 250 V/m)
±1 dB (from 1 μW/cm² to 16.5 mW/cm²)
±2 dB (250 to 400 V/m)
±2 dB (from 16.5 mW/cm² to 42.44 mW/cm²)
±2 dB (from 16.5 mW/cm² to 170 mW/cm²)
±3 dB (0.265 μW/cm² to 1 μW/cm²)
±3 dB (1 to 2 V/m)
Izotropinis nuokrypis
±1.25 dB (100 MHz to 10 GHz)
±2 dB (more than 10 GHz to 26.5 GHz)
±2 dB typ. (more than 26.5 GHz)
Temperatūros reakcija
±0.2 dB (-20 °C to +50 °C, related to 23 °C)
Akredituotas kalibravimas
DAkkS, ILAC-MRA (DIN EN ISO/IEC 17025, IEEE Std. 1309)
For measurands outside the scope, a factory calibration is performed.
Rekomenduojamas kalibravimo intervalas
24 mėn
Priedai
Atsisiųsti
FieldMan Brochure

FieldMan Brochure, v. NSTS 03/23 E0384B, English

Atsisiųsti (6.51MB)
FieldMan Sondes

FieldMan - Probes program for measurements up to 90 GHz

Atsisiųsti (100.62KB)
Datasheet Probe EFD-9091

Datasheet Probe EFD-9091, v. NSTS 0523-E0405A, English

Atsisiųsti (220.38KB)