EFD-5091 FieldMan E-Field zonde, 300MHz - 50GHz
  • EFD-5091 FieldMan E-Field zonde, 300MHz - 50GHz

EFD-5091 FieldMan E-Field zonde, 300MHz - 50GHz

EFD-5091

Isotropic measurement of electric fields from 300 MHz to 50 GHz using the Field Meter FieldMan®

Apraksts

The probe contains three orthogonally-arranged thermo-couple sensors and detects electric fields from 300 MHz to 50 GHz, covering the fields found in telecommunications, satellite communications and radar equipment. The use of thermocouples naturally results in a true RMS reading, making the probe particularly suitable for measuring human safety limit values in a multi-frequency environment.

The probe's interface digitally transmits the measurement data to the base unit, which has no individual influence on the measured values and therefore does not need to be calibrated. The accredited probe calibration is carried out at several frequencies. The calibration data is stored in the probe and is automatically taken into account during the measurement. If the frequency of the predominating field strength is known, a correction factor can additionally be applied to increase the measurement accuracy.

  • Measurement of field exposure from telecommunications, satellite communications and radar in working environments
  • True RMS value display even when several, strongly pulsed signals are superimposed
  • Isotropic (non-directional) measurement
  • Digital probe interface - no more meter calibration
  • Self-test of the probe interface with integrated sensor function test
  • Automatic offset correction, no zero adjustment required
  • Lowest temperature response due to thermocouple sensor technology and wide temperature range down to -20 °C
  • High Immunity at 50/60 Hz
  • Accredited calibration included
Produkta detaļas
EFD-5091

Datu lapa

Lauks
Elektriskais lauks
Izmēri
308 mm x 66 mm Ø
Svars
up to 100 g
Frekvenču diapazons
300MHz - 50GHz
Ekspluatācijas apstākļi
-20 °C to +50 °C; up to 29 g/m³ (up to 93 % RH at +30 °C), non-condensing
Uzglabāšanas apstākļi
1K5 (IEC 60721-3) -40 °C to +70 °C
Aizsardzības klase
IP54 (probe screwed on)
Produkta veids
FieldMan probes
Dinamiskais diapazons (dB)
37 dB
Mērījumu diapazons, E-lauks
8 to 614 V/m (True RMS); 17 μW/cm² to 100 mW/cm² (True RMS)
Integrācijas laiks
270 ms
Type of frequency response
Flat
Nepārtraukta viļņa bojājuma līmenis
1200 V/m; 0.4 W/cm²
Maksimālais bojājums
20 kV/m; 100 W/cm²
Sensora tips
Thermocouple
Virzība
Izotrops (trīsass)
Telpiskais novērtējums
Combined 3-axis (RSS)
Paraugu ņemšanas ātrums
5 Hz
Temperatūras sensori
Integrated sensors for displaying the ambient temperature and for automatic offset compensation
Pašpārbaude
Interface function test and sensor test for interruption of the sensors
Frekvences reakcijas vienmērīgums
±1.5 dB (1.8 GHz to 26.5 GHz); +1.5/-2 dB (up to 26.5 GHz to 40 GHz)
Linearitātes novirze
±1 dB (19.5 to 61.4 V/m); ±0.3 dB (less than 61.4 V/m); ±1 dB (0.1 to 1 mW/cm²) ±0.3 dB (less than 1 mW/cm²)
Izotropiskā novirze
±1 dB (0.6 GHz to 10 GHz); ±1.2 dB (10 GHz to 45.5 GHz)
Temperatūras reakcija
±0 dB
Akreditēta kalibrēšana
DAkkS, ILAC-MRA (DIN EN ISO/IEC 17025, IEEE Std. 1309); For measurands outside the scope, a factory calibration is performed.
Ieteicamais kalibrēšanas intervāls
24 months
Pielikumi
Lejupielādēt
FieldMan Brochure

FieldMan Brochure, v. NSTS 03/23 E0384B, English

Lejupielādēt (6.51MB)
Datasheet Probe EFD-5091

Datasheet Probe EFD-5091, v. NSTS 1122-E0389A, English

Lejupielādēt (308.76KB)