EFD-0391 FieldMan E-Field probe, 100kHz - 6GHz
  • EFD-0391 FieldMan E-Field probe, 100kHz - 6GHz

EFD-0391 FieldMan E-Field probe, 100kHz - 3GHz

EFD-0391

Isotropic measurement of electric fields from 100kHz to 6GHz using the Field Meter FieldMan®

Apraksts

The probe detects electric fields from 100 kHz to 3 GHz, covering the fields that occur in broadcasting, telecoms, and industry. The high sensitivity of 0.2 V/m and excellent linearity make it ideal for measuring human safety limit values in the general public domain.

The probe's interface digitally transmits the measurement data to the base unit, which has no individual influence on the measured values and therefore does not need to be calibrated. The accredited probe calibration is carried out at several frequencies. The calibration data is stored in the probe and is automatically taken into account during the measurement. If the frequency of the predominating field strength is known, a correction factor can additionally be applied to increase the measurement accuracy.

  • General public and occupational field exposure from broadcasting, telecoms and industrial equipment
  • Isotropic (non-directional) measurement
  • 64 dB dynamic without changing range
  • Excellent sensitivity starting at 0.2 V/m
  • Digital probe interface - no more meter calibration
  • Self-test of the probe interface with integrated sensor function test
  • Automatic offset correction, no zero adjustment required
  • Temperature compensation of the sensors for minimum drift and wide temperature range down to -20 °C
  • High Immunity at 50/60 Hz
  • Accredited calibration included
Produkta detaļas
EFD-0391

Datu lapa

Lauks
Elektriskais lauks
Izmēri
308 mm x 66 mm Ø
Svars
up to 100 g
Frekvenču diapazons
100kHz - 3GHz
Ekspluatācijas apstākļi
-20 °C to +50 °C; up to 29 g/m³ (up to 93 % RH at +30 °C), non-condensing
Uzglabāšanas apstākļi
1K5 (IEC 60721-3) -40 °C to +70 °C
Aizsardzības klase
IP54 (probe screwed on)
Produkta veids
FieldMan probes
Dinamiskais diapazons (dB)
64dB
Mērījumu diapazons, E-lauks
0.2 to 320 V/m (CW), 0.2 to 10 V/m (True RMS), 10 nW/cm² to 27 mW/cm² (CW), 10 nW/cm² to 0.027 mW/cm² (True RMS)
Integrācijas laiks
265 ms
Type of frequency response
Flat
Nepārtraukta viļņa bojājuma līmenis
800 V/m, 170 mW/cm²
Maksimālais bojājums
8 kV/m, 17 W/cm²
Sensora tips
Diode based system
Virzība
Izotrops (trīsass)
Telpiskais novērtējums
3 separate axes
Paraugu ņemšanas ātrums
5 Hz
Temperatūras sensori
Integrated sensors for displaying the ambient temperature, for automatic offset compensation and for compensation of the temperature response
Pašpārbaude
Interface function test and sensor test for short circuit and interruption of diodes
Frekvences reakcijas vienmērīgums
±1 dB (1 MHz to 1 GHz); ±1.25 dB (1 GHz to 2.45 GHz)
Linearitātes novirze
±0.5 dB (1.2 to 200 V/m); ±0.7 dB (200 to 320 V/m); ±0.5 dB (0.00038 to 10.6 mW/cm²);±0.7 dB (10.6 to 27 mW/cm²)
Izotropiskā novirze
±1 dB
Temperatūras reakcija
+0.1/ -0.8 dB (0 °C to 50 °C, related to 23 °C); +1.5/ -0.8 dB (-20 °C to +50 °C, related to 23 °C)
Akreditēta kalibrēšana
DAkkS, ILAC-MRA (DIN EN ISO/IEC 17025, IEEE Std. 1309); For measurands outside the scope, a factory calibration is performed.
Ieteicamais kalibrēšanas intervāls
24 months
Pielikumi
Lejupielādēt
FieldMan Brochure

FieldMan Brochure, v. NSTS 03/23 E0384B, English

Lejupielādēt (6.51MB)
Datasheet Probe EFD-0391

Datasheet Probe EFD-0391, v. NSTS 1122-E0385A, English

Lejupielādēt (311.48KB)
Komentāri (0)
Novērtējums
Šobrid vēl nav klientu atsauksmju.
10 produktu tajā pat kategorijā:
Jums arī varētu patikt