Apraksts

The probe detects electric fields from 2 MHz up to 40 GHz. This area covers the frequencies used for broadcasting, as well as applications for telecommunications, satellite communications and radar technologies. The probe’s high linearity and sensitivity make it ideal for checking human safety limit values in the general public and occupational environments. 

The probe's interface digitally transmits the measurement data to the base unit, which has no individual influence on the measured values and therefore does not need to be calibrated. Probe calibration is carried out at several frequencies and is accredited up to 18 GHz. The calibration data is stored in the probe and is automatically taken into account during the measurement. If the frequency of the predominating field strength is known, a correction factor can additionally be applied to increase the measurement accuracy.

  • Isotropic (non-directional) measurement
  • Excellent sensitivity starting at 1.2 V/m
  • Wide True RMS range up to 65 V/m
  • Digital probe interface - no more meter calibration
  • Self-test of the probe interface with integrated sensor function test
  • Automatic offset correction, no zero adjustment required
  • Wide temperature range from -20 °C to +50 °C
  • Calibration is performed up to 40 GHz and is accredited up to 18 GHz
Produkta detaļas
EFD-4091

Datu lapa

Lauks
Elektriskais lauks
Izmēri
308 mm x 66 mm Ø
Svars
up to 100 g
Frekvenču diapazons
2MHz - 40GHz
Ekspluatācijas apstākļi
-20 °C to +50 °C
up to 29 g/m³ (up to 93 % RH at +30 °C), non-condensing
Aizsardzības klase
IP54 (probe screwed on)
Produkta veids
FieldMan probes
Dinamiskais diapazons (dB)
56.5
Mērījumu diapazons, E-lauks
1.2 to 65 V/m (True RMS), 0.382 μW/cm² to 1.0 mW/cm² (True RMS)
1.2 to 800 V/m (CW), 0.382 μW/cm² to 170 mW/cm² (CW)
Type of frequency response
Flat
Nepārtraukta viļņa bojājuma līmenis
1500 V/m, 600 mW/cm²
Maksimālais bojājums
1800 V/m, 860 mW/cm²
Sensora tips
Diode based system
Virzība
Izotrops (trīsass)
Telpiskais novērtējums
Combined 3-axis (RSS)
Paraugu ņemšanas ātrums
5 Hz / 265 ms
Temperatūras sensori
Integrated sensors for displaying the ambient temperature, for automatic offset compensation and for compensation of the temperature response
Pašpārbaude
Interface function test and sensor test for interruption of diodes
Frekvences reakcijas vienmērīgums
typ. -3dB @ 2 MHz / 40 GHz
±1.5 dB (3 MHz to 15 GHz)
±3 dB (from 15 GHz to 35 GHz)
Linearitātes novirze
±1 dB (1 μW/cm² to 16.5 mW/cm²)
±1 dB (2 to 250 V/m)
±1 dB (from 1 μW/cm² to 16.5 mW/cm²)
±2 dB (from 16.5 mW/cm² to 170 mW/cm²)
±2 dB (from 250 to 800 V/m)
±3 dB (0.38 to 1 μW/cm²)
±3 dB (1 to 2 V/m)
Izotropiskā novirze
typ. ±2 dB @ 2 MHz
typ. ±6 dB @ 40 GHz
±1.0 dB (10 MHz to 5 GHz)
±1.5 dB (from 5 GHz to 20 GHz)
±2 dB (more than 10 GHz to 26.5 GHz)
±3.0 dB (from 20 GHz to 30 GHz)
Temperatūras reakcija
+0.5/ -0.5 dB (0 °C to 50 °C, related to 23 °C)
+1.5/ -1.5 dB (-20 °C to +50 °C, related to 23 °C)
Akreditēta kalibrēšana
DAkkS, ILAC-MRA (DIN EN ISO/IEC 17025, IEEE Std. 1309)
For measurands outside the scope, a factory calibration is performed.
Ieteicamais kalibrēšanas intervāls
24 mēneši
Pielikumi
Lejupielādēt
FieldMan Brochure

FieldMan Brochure, v. NSTS 03/23 E0384B, English

Lejupielādēt (6.51MB)
FieldMan Sondes

FieldMan - Probes program for measurements up to 90 GHz

Lejupielādēt (100.62KB)
Datasheet Probe EFD-4091

Datasheet Probe EFD-4091, v. NSTS 0523-E0399A, English

Lejupielādēt (229.55KB)
Komentāri (0)
Novērtējums
Šobrid vēl nav klientu atsauksmju.
10 produktu tajā pat kategorijā:
Jums arī varētu patikt